Mass Flow Meter – 0-300 LPM – Air/O2/N2

Mass Flow Meter – 0-300 LPM – Air/O2/N2 – 22mm ISO

The TSI Mass Flowmeter 4040 can be used for a multitude of gas flow measurement applications. Whether measuring gas flows in a laboratory or manufacturing setting, TSI general purpose mass flowmeters provide accurate results with multiple data output options. The display versions come complete with accessories making setup and operation fast and convenient.

The TSI mass flowmeters are specified as two percent of flow rate reading. One TSI mass flow meter covers the same range as three percent (or more) of full scale devices, with better accuracy at most points. Fast four millisecond response ensures accuracy in fluctuating flows. This fast response time of the TSI mass flowmeter is ideal for closed-loop control systems and integrated volume measurements. Low pressure drop of the TSI mass flowmeter minimizes flow circuit back pressure and its impact on the system.

SKU: 751 Category:

FEATURES AND BENEFITS

Fast and accurate
Low pressure drop

APPLICATIONS

Medical
Industrial
Laboratory

Base not included.

FAQ
ARE THE FLOWMETERS BI-DIRECTIONAL?
The flowmeters cannot detect the direction of the flow only the magnitude of the flow. Also the meters are calibrated in the direction indicated by the arrow on the bottom of the flowmeter. See the application note Reverse Flow Effects on TSI General Purpose Flowmeters for the correction for reverse flows.
ARE THERE ANY SPECIAL CONNECTION REQUIREMENTS FOR THE ANALOG OUTPUT?
Yes. The analog signal ground return is connected to the power ground inside the flowmeter. For this reason your equipment must use a voltmeter or analog input circuit with a differential input or floating ground. Failure to use this type of ground will result in measurement errors.
ARE THERE CORRECTION FACTORS FOR THESE DIFFERENT GASSES?
No. These flowmeters are separately calibrated for air and 100% O2. Calibrations for the other gasses involves an internal algorithm in the meter. The desired gas is selected by an RS232 command. Once set, the LCD display indicates the gas calibration. The flow rate displayed on the LCD and the flowmeter?s outputs is then automatically corrected for the selected gas.
ARE THESE FLOWMETERS PRESSURE COMPENSATED?
Yes. The output readings are compensated for pressure changes as long as those pressures changes are within 0.5 to 2.0 atmospheres absolute pressure.
ARE THESE FLOWMETERS TEMPERATURE COMPENSATED?
Yes. The output readings are compensated for temperatures changes as long as those temperature changes are within the temperature-compensated range.